Chino, EL3000 Series, Analog Recorders, 100mm, Dot Printing Type, Recorders

Chino EL3000 Series Analog Recorders
100mm (Dot Printing Type)


EL3000 Series Analog Recorders 100mm (Dot Printing Type)

EL3000 series is a dot printing type analog recorder sized 144x144mm with 100mm width chart. The unit starts recording as soon as the power supply and input are connected and it is also easy to operate. Scale plate, input range and function of the recorder can be selected for various purpose and applications as many kinds of options are prepared. EL3000 Series
Universal power supply
Universal power supply with voltage range of 100 to 240V AC (50/60Hz) is applied.
Linear temperature scale
The accuracy is ±0.05% and the resolution is 1ƒÊV or 0.1°C
Various industrial values can be measured at the same time with selectable ranges
Temperature scale of thermocouple and resistance thermometer input is a linear scale that is excellent in reading indication value.
Standard 6 chart speeds
6 chart speeds (5,10,20,40,80,160mm/h) are switchable as standard. 5 chart speed and hour/minute change are prepared as option.
Alarm setting (common alarm) as standard
Higher and lower limit alarm can be programmed for every point. Alarm value is easy to be programmed by pointer location. You can check the alarm by front LED lighting. Alarm output is prepared as option.
CE marking
The EL recorder is conformed to the rules of safety standards of CE. (Conformity pending)
Unit structure and light-weight
Light-weight (50% of the previous unit weight) was realized by easy maintenance unit structure.
Employing removable type terminal board
Employing easy connecting removable type terminal board.

Multi-Point Type
Thermocouple
Model Specification Remarks
EL3D15-NNN 1 point Accessories
EL3D25-NNN 2 points Accessories
EL3D35-NNN 3 points Accessories
EL3D65-NNN 6 points Accessories
RTD *1
EL3D17-NNN 1 point Accessories
EL3D27-NNN 2 points Accessories
EL3D37-NNN 3 points Accessories
EL3D67-NNN 6 points Accessories
*1 Optional built-in voltage divider and thermocouple burnout input is only type "7"

Option
Model Specification
EL3D** - NN w/Input and scale plate

: 0: Standard input + standard scale plate
1: Non-standard input (including current input and built-in voltage divider) + standard scale
2: Standard input + non-standard scale plate
3: Non-standard input (including current input and built-in voltage divider) + Non-standard scale
*2 Double scale is available.
Input and scale selection are needed for non-standard input and non-standard scale plate.

Model Specification
EL3D** - Alarm
: 0: None
1: 2 alarm outputs

Model Specification
EL3D*7 - Chart speed + burnout
  0: Standard 6-speed + burnout disabled
1: Standard 6-speed + up-scale burnout
2: Standard 6-speed + down-scale burnout
A: Standard 5-speed hour/minute change + burnout disabled
B: Standard 5-speed hour/minute change + up-scale burnout
C: Standard 5-speed hour/minute change + down-scale burnout
*3 Burnout on all channels is programmed together for thermocouple/residence thermometer input.

Input Specifications
Measuring points: 1,2,3 and 6 points
Reference range and types: DC voltage: ±13.8mV, ±27.6mV, ±69mV, 200mV, ±500mV,±2V, ±5V Built-in voltage divider; ±10V, ±25V, ±50VD
Ccurrent: External installation of shunt resistor(250Ω) is applied (option)
Thermocouples: K, E, J, T, R, and B (option)
Resistance thermometer: Pt100(1997) (Measured current; 1mA)* Linear scale for thermocouple and resistance thermometer
Input designation: Single scale (standard), double scale (option)
Accuracy rating: ±0.5% of input span (except for some input under standard operating condition) Refer to the table of standard range and minimum width of scale for non-standard input
Indicating deadband: 0.3% of input span
Reference junction compensation accuracy: K,E,J,T --- ±1.0°C or less (23°C±10°C) ±2.0°C or less (0 to 50°C)
(For internal reference junction compensation, the errors above are added to the accuracy rating)
Temperature drift: ±0.02%/°C (Converted into reference ranges)
Measurement cycle: 6 seconds/point
Indicating resolution: Approximately 1/2,000
Burnout (option): On thermocouple or resistance thermometer input, disconnection of signal can be detected
.(Up-scale and down scale burnout on all channels can be programmed.)
Burnout detection: Voltage application method (approximately 8V, 1mA)
Allowable signal source resistance: Thermocouple inputs, DC voltage inputs (±5V or less) --- 1kΩ(burnout disabled) or less
DC Voltage inputs (input more than ±5V) --- 100Ω or less
Resistance thermometer inputs --- per wire 10Ω or less (Same resistance for 3 wires)
Maximum input voltage: Thermocouple inputs, DC voltage inputs --- ±10V DC or less
DC voltage inputs (Voltage divider built-in) --- ±60V DC or less
Resistance thermometer --- ±6V DC or less
Maximum common mode voltage: 30V AC
Common mode rejection ratio: 120dB or more (50/60Hz±0.1%)
Normal mode rejection ratio: 50dB or more (50/60Hz±0.1%)
Recording Specifications
Recording accuracy: ±0.5% of recording span
Recording system: Inkpad dotting
Balancing time: Input span movement --- approximately 2 seconds
Recording color: 1: red 2: blue 3: green 4: violet 5: purple 6:brown
Chart paper: Fan-fold type: total width of 114mm, total length of 10m, effective chart width of 100mm
Chart speed: 6-speed change, 5,10,20,40,80,160mm/h (standard)
Chart speed accuracy: ±0.1% or less (It is based on the chart scale.)
Indicating Specifications
Analog indication: Scale plate and pointer
Scale plate: Single scale or double scale (minimum scale division: 80)
Alarm Specifications
Alarm display: Pointer and alarm-point seal pasted on scale.
Alarm LED lamp lightens for alarming
(All channels OR output)
Alarm types: Higher and lower-limit alarm
Alarm programming: Individual setting for higher and lower-limit value
(Programming percentage of input span by indicating pointer, input resolution 0.5%)
Alarm deadband: 0.4% of input span
Alarm output (option): 1a contact and 2 outputs (common)
Maximum contact capacity: 2A (resistive load), 0.5A (inductive load)
General Specifications
Rated power voltage: 100 to 240V AC, 50/60Hz (Universal power supply) with power supply switch
Power consumption: Maximum 12VA (100V AC), Maximum 20VA (240V AC)
Environmental conditions: Reference operation condition
Ambient temperature range: 21 to 25°C Ambient humidity range: 45 to 65%RH Power voltage: 100V AC±1% Power frequency: 50/60Hz±0.5% Attitude: left/right 0°, forward tilting 0°, backward tilting 0° Warm-up time: longer than 30 minutes

Normal operation condition
Ambient temperature range: 0 to 50°C Ambient humidity range: 20 to 80%RH Power voltage: 90 to 264V AC Power frequency: 50/60Hz ±2% Attitude: left/right 0 to 10°, forward tilting 0°, backward tilting 0 to 20°

Transportation condition (at the packed condition on shipment from our factory)
Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation) Vibration: 10 to 60Hz, 4.9m/ S2 (0.5G) or less Impact: 392m/S2 (40G) or less

Storage condition
Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation)
Insulation resistance: Secondary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary and secondary terminals --- 20MΩ or more at 500V DC
Note:Primary terminals --- Power (L,N), Alarm terminals (mechanical relay)
Secondary terminals --- Measurement input terminals
Dielectric strength: Secondary terminals and protective conductor terminals --- 1 minute at 500V AC
Primary terminals and protective conductor terminals --- 1 minute at 1500V AC
Primary and secondary terminals --- 1 minute at 2300V AC Notes; Primary terminals --- Power(L,N), Alarm terminals (mechanical relay)
Secondary terminals --- Measurement input terminals
Case: Door (frame) --- ABS resin, window --- glass Case --- ABS resin
Color: Door (frame) --- Black (equivalent to Mussel N1.5),window --- Transparent Case --- Black (equivalent to Mussel N1.5)
Mounting: Panel mounting
Weight: Approximately 1.7kg
Power voltage fluctuation: Indication fluctuation 0.2% or less (conversed into reference ranges at 90 to 264V AC)
Optional Specifications
Alarm output Alarm contact output is available
Alarm relay --- Mechanical relay 1a contact, 2 outputs (common)
Maximum contact rating: 250V AC 2A, 30V DC 2A (resistive load), 250V AC 0.5A, 30V DC 0.5A (inductive load)
DC current input 250Ω of shunt resistor is applied to measure voltage input
Built-in voltage divider Built-in voltage divider(1/1000) measures input in the range of ±5V to ±50V (input type "7" only)
Non-standard input Refer to the table of standard range and programmable minimum width of scale Minimum width of scale ---
DC voltage: 10mV DC width or mor
Thermocouple: K; 250°C width or moreE,J,T; 200°C width or moreR; 800°C width or more
Resistance thermometer: 100°C width or more
Double scale Measures input with 2 types of scales (each scale is only serial channel)
Burnout Function for detecting disconnection for sensor with thermocouple or resistance thermometer input.
Up-scale and down scale burnout on all channels can be programmed (Input type "7" only), parallel operation is not possible
Chart speed Hour/minute change 5-speed change, 5,10,20,40,80mm/minute, hour change
16m chart paper Maximum length 15.6m

EL3000 Series Electronic Recorders
Items Recorder Type Part Number Qty
EL3000 Ink Pad Case EL3000/ES600
1 Point Dotting
22023-039001
81-0056
1 Box (5 ea.)
EL3000 Ink Pad Case EL3000/ES600
2 Points Dotting
22023-039002
81-0067
1 Box (5 ea.)
EL3000 Ink Pad Case EL3000/ES600
3 Points Dotting
22023-039003
81-0078
1 Box (5 ea.)
EL3000 Ink Pad Case EL3000/ES600
6 Points Dotting
22023-039006
81-0089
1 Box (5 ea.)
Chart Paper - 100mm (Standard) All Type EM001 1 Box (15 sheets)
PS Sheet EL3000 Series (1 Pen Type) PSE-355 Download

Quality Temperature, Pressure and Recording Instrumentation